Device recall: Nextgen Complete Knee Solution Femoral and Provisional Impactor/ Extractor, REF 00-5901-026-00, Non-Sterile Zimmer…
Class II FDA enforcement event 61764. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Zimmer Inc. started this device recall on 30 Apr 2012. The firm address in the FDA record is in Warsaw, IN. FDA put the recall in Class II. FDA event 61764 has 2 product records. The first FDA enforcement report date is 8 Aug 2012. The record gives the status Terminated, with the termination date 4 Jun 2014.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution - USA (nationwide) and the countries of Singapore, China, Hong Kong, Canada, Australia, Japan, Europe, Germany, India, and Finland.
Recalling firm
Zimmer Inc., Warsaw, IN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2091-2012 | Nextgen Complete Knee Solution Femoral and Provisional Impactor/ Extractor, REF 00-5901-026-00, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components. | 1662 units | Class II | Terminated | 8 Aug 2012 |
| Z-2092-2012 | Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-40, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components. | 65 units | Class II | Terminated | 8 Aug 2012 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2012
- 823
- Share of these events in Class II
- 88%
- Product records in this event
- 2Median of the firm: 1
- Events of the firm on the register
- 108
Other events of the firm
Official channels
- FDA enforcement report Z-2091-2012The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 61764. https://recallregister.com/fda/ii/61764/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution - USA (nationwide) and the countries of Singapore, China, Hong Kong, Canada, Australia, Japan, Europe, Germany, India, and Finland.
- What quantity of product is in FDA event 61764?
- The event has 2 product records. The first record gives this quantity:
1662 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2012 have the same class?