Recall Register

Device recall: System 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the…

Class II FDA enforcement event 61242. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Terumo Cardiovascular Systems (Terumo CVS) received eight complaints since September 2008 of drive motor failure of the Sarns Centrifugal System of the Terumo Advanced Perfusion System 1. A failure of the drive motor for the Terumo System 1 centrifugal system triggers error messages and alarms; therefore it is easily recognizable by the user. The potential hazards are dependent on the application

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Terumo Cardiovascular Systems Corporation started this device recall on 22 Jun 2012. The firm address in the FDA record is in Ann Arbor, MI. FDA put the recall in Class II. FDA event 61242 has 2 product records. The first FDA enforcement report date is 1 Aug 2012. The record gives the status Terminated, with the termination date 6 Mar 2013.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA (nationwide) and the countries of Nationwide and worldwide: ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala,…

Recalling firm

Terumo Cardiovascular Systems Corporation, Ann Arbor, MI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 61242: product records
Recall numberProductQuantityClassStatusReport date
Z-2055-2012System 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this…992 devicesClass IITerminated1 Aug 2012
Z-2056-2012System 1 Base 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this…640 devicesClass IITerminated1 Aug 2012

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class II
88%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
73

Other events of the firm

Other FDA enforcement events of Terumo Cardiovascular Systems Corporation
InitiatedProduct (first record)TypeClassRecords
28 May 2025CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial KDeviceClass II1
3 Mar 2025Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754DeviceClass II1
20 Feb 2024Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312 6DeviceClass II1
1 Mar 2023CAPIOX RX15 Hollow Fiber Oxygenator, Catalog Number 3CX*RX15RE30DeviceClass II3
8 Jun 2021Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188DeviceClass II1
16 Apr 2021The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directlDeviceClass II3
2 Mar 2021Consist of a distal connection mechanism (either standard Terumo Ball Quick Connect or Terumo Shaft Quick Connect), an aDeviceClass II1
11 Nov 2020(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, Sterile Catalog NuDeviceClass II1
25 Aug 2020Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for useDeviceClass II3
10 Jan 2020Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for DeviceClass II4

Terumo Cardiovascular Systems Corporation: all 73 events

Official channels

Cite this page

Recall Register. FDA enforcement event 61242. https://recallregister.com/fda/ii/61242/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA (nationwide) and the countries of Nationwide and worldwide: ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala,…
What quantity of product is in FDA event 61242?
The event has 2 product records. The first record gives this quantity: 992 devices
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
88% of the 823 device events that started in 2012 are in Class II.