Recall Register

Device recall: Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the…

Class II FDA enforcement event 53337. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medacta Usa Inc started this device recall on 13 Oct 2008. The firm address in the FDA record is in Camarillo, CA. FDA put the recall in Class II. FDA event 53337 has 1 product record. The first FDA enforcement report date is 14 Nov 2012. The record gives the status Terminated, with the termination date 9 Nov 2012.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.

Recalling firm

Medacta Usa Inc, Camarillo, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 53337: product records
Recall numberProductQuantityClassStatusReport date
Z-0267-2013Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.12 unitsClass IITerminated14 Nov 2012

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2008
18
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
12

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 53337. https://recallregister.com/fda/ii/53337/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.
What quantity of product is in FDA event 53337?
The FDA record gives this quantity: 12 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2008 have the same class?
89% of the 18 device events that started in 2008 are in Class II.