Recall Register

Device recall: Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS.…

Class II FDA enforcement event 49516. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Ventana Medical Systems Inc started this device recall on 12 Jun 2008. The firm address in the FDA record is in Tucson, AZ. FDA put the recall in Class II. FDA event 49516 has 1 product record. The first FDA enforcement report date is 4 Sep 2013. The record gives the status Terminated, with the termination date 29 Aug 2013.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA Nationwide and the countries of Canada, Europe, Asia, Japan.

Recalling firm

Ventana Medical Systems Inc, Tucson, AZ.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 49516: product records
Recall numberProductQuantityClassStatusReport date
Z-2118-2013Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The Benchmark product lines are intended to automatically stain histological or cytological specimens on microscope slides with specific immunohistochemistry or in situ…1217 units totalClass IITerminated4 Sep 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2008
18
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
10

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 49516. https://recallregister.com/fda/ii/49516/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA Nationwide and the countries of Canada, Europe, Asia, Japan.
What quantity of product is in FDA event 49516?
The FDA record gives this quantity: 1217 units total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2008 have the same class?
89% of the 18 device events that started in 2008 are in Class II.