Recall Register

Device recall: VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.

Class I FDA enforcement event 99835. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Ventec Life Systems, Inc. started this device recall on 17 Sep 2026. The firm address in the FDA record is in Bothell, WA. FDA put the recall in Class I. FDA event 99835 has 12 product records. The first FDA enforcement report date is 28 Oct 2026. The record gives the status Ongoing. 33 of the 697 device events that started in 2026 have more product records. The median device event of 2026 has 1 product record.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Distribution US nationwide including Puerto Rico.

Recalling firm

Ventec Life Systems, Inc., Bothell, WA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

12 product records in the event.

FDA event 99835: product records
Recall numberProductQuantityClassStatusReport date
Z-0122-2027VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.4,369 unitsClass IOngoing28 Oct 2026
Z-0123-2027VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.325 unitsClass IOngoing28 Oct 2026
Z-0124-2027VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support.1,534 unitsClass IOngoing28 Oct 2026
Z-0125-2027VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.606 unitsClass IOngoing28 Oct 2026
Z-0126-2027VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.1,532 unitsClass IOngoing28 Oct 2026
Z-0127-2027VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.100 unitsClass IOngoing28 Oct 2026
Z-0128-2027VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.4,368 unitsClass IOngoing28 Oct 2026
Z-0129-2027V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.39,568 unitsClass IOngoing28 Oct 2026
Z-0130-2027VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support.930 unitsClass IOngoing28 Oct 2026
Z-0131-2027VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.180 unitsClass IOngoing28 Oct 2026
Z-0132-2027V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support.30,120 unitsClass IOngoing28 Oct 2026
Z-0133-2027V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.25,112 unitsClass IOngoing28 Oct 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2026
697
Share of these events in Class I
11%
Product records in this event
12Median of the firm: 1
Events of the firm on the register
7

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 99835. https://recallregister.com/fda/i/99835/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
3 Nov 2026
Newest FDA report date on the register
28 Oct 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Distribution US nationwide including Puerto Rico.
What quantity of product is in FDA event 99835?
The event has 12 product records. The first record gives this quantity: 4,369 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
11% of the 697 device events that started in 2026 are in Class I.