Recall Register

Device recall: Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK…

Class I FDA enforcement event 98720. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

American Contract Systems Inc started this device recall on 19 Mar 2026. The firm address in the FDA record is in Tiffin, OH. FDA put the recall in Class I. FDA event 98720 has 1 product record. The first FDA enforcement report date is 13 May 2026. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US distribution to MN. No OUS distribution.

Recalling firm

American Contract Systems Inc, Tiffin, OH.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 98720: product records
Recall numberProductQuantityClassStatusReport date
Z-1981-2026Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ7,311 kitsClass IOngoing13 May 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2026
572
Share of these events in Class I
12%
Product records in this event
1Median of the firm: 4
Events of the firm on the register
22

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 98720. https://recallregister.com/fda/i/98720/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US distribution to MN. No OUS distribution.
What quantity of product is in FDA event 98720?
The FDA record gives this quantity: 7,311 kits
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
12% of the 572 device events that started in 2026 are in Class I.