Recall Register

Device recall: Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7

Class I FDA enforcement event 98586. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode…

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Insulet Corporation started this device recall on 12 Mar 2026. The firm address in the FDA record is in Acton, MA. FDA put the recall in Class I. FDA event 98586 has 1 product record. The first FDA enforcement report date is 22 Apr 2026. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide distribution.

Recalling firm

Insulet Corporation, Acton, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 98586: product records
Recall numberProductQuantityClassStatusReport date
Z-1797-2026Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G71,240,115 unitsClass IOngoing22 Apr 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2026
572
Share of these events in Class I
12%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
15

Other events of the firm

Other FDA enforcement events of Insulet Corporation
InitiatedProduct (first record)TypeClassRecords
20 May 2026Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-PackDeviceClass I3
11 Sep 2025The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnifyDeviceClass II1
23 May 2025Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all PDeviceClass II1
30 Nov 2023Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 AppDeviceClass I1
28 Feb 2023Omnipod 5 Automated Insulin Delivery SystemDeviceClass II1
14 Nov 2022Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-00042DeviceClass II1
17 Oct 202218239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: AssemblDeviceClass I1
1 Feb 2021Omnipod DASH Insulin Management System User Guide (OUS Only) Product Catalog Number: PT 000002 XXX XXX XX PT 000318 PT 0DeviceClass II1
2 Apr 2020Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number PT000010 M/D: INT1D001MG - Product Usage: iDeviceClass II2
13 Feb 2020Omnipod DASH Personal Diabetes Manager (PDM), Catalog numbers US: 18239 M/D: PDM-USA1-D001-MG-USA1 EU: PT-000010 - M/D: DeviceClass II1

Insulet Corporation: all 15 events

Official channels

Cite this page

Recall Register. FDA enforcement event 98586. https://recallregister.com/fda/i/98586/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide distribution.
What quantity of product is in FDA event 98586?
The FDA record gives this quantity: 1,240,115 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
12% of the 572 device events that started in 2026 are in Class I.