Recall Register

Device recall: Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ…

Class I FDA enforcement event 98277. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Medline Industries, LP started this device recall on 22 Dec 2025. The firm address in the FDA record is in Northfield, IL. FDA put the recall in Class I. FDA event 98277 has 7 product records. The first FDA enforcement report date is 18 Feb 2026. The record gives the status Ongoing. 70 of the 945 device events that started in 2025 have more product records. The median device event of 2025 has 1 product record. The openFDA file has 945 device events that started in 2025 and 972 that started in 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Distribution US nationwide.

Recalling firm

Medline Industries, LP, Northfield, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

7 product records in the event.

FDA event 98277: product records
Recall numberProductQuantityClassStatusReport date
Z-1319-2026Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;228Class IOngoing18 Feb 2026
Z-1320-2026Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10…311Class IOngoing18 Feb 2026
Z-1321-2026Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable…41Class IOngoing18 Feb 2026
Z-1322-2026Reprocessed St. Jude Supreme and Response Diagnostic Electrophysiology Catheters: St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 2-5-2mm 401211RH 401222RH 401223RH 401226RH 401227RH 401260RH 401261RH 401306RH 401309RH 401312RH 401430RH 401433RH 401435RH 401441RH 401442RH 401443RH 401448RH…48Class IOngoing18 Feb 2026
Z-1323-2026Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554¿650Class IOngoing18 Feb 2026
Z-1324-2026Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on GE Systems 10439072RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on Siemens Systems 10438577RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound…94Class IOngoing18 Feb 2026
Z-1325-2026Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH511Class IOngoing18 Feb 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class I
13%
Product records in this event
7Median of the firm: 5
Events of the firm on the register
74

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 98277. https://recallregister.com/fda/i/98277/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Distribution US nationwide.
What quantity of product is in FDA event 98277?
The event has 7 product records. The first record gives this quantity: 228
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?
13% of the 945 device events that started in 2025 are in Class I.