Recall Register

Device recall: AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;

Class I FDA enforcement event 98237. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional…

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Boston Scientific Corporation started this device recall on 19 Dec 2025. The firm address in the FDA record is in Marlborough, MA. FDA put the recall in Class I. FDA event 98237 has 6 product records. The first FDA enforcement report date is 11 Feb 2026. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France,…

Recalling firm

Boston Scientific Corporation, Marlborough, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 98237: product records
Recall numberProductQuantityClassStatusReport date
Z-1158-2026AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;557 unitsClass IOngoing11 Feb 2026
Z-1159-2026AXIOS Stent and Electrocautery-Enhanced Delivery System - 6mm x 8mm; Material Number (UPN): M00553680;39 unitsClass IOngoing11 Feb 2026
Z-1160-2026AXIOS Stent and Electrocautery-Enhanced Delivery System - 8mm x 8mm; Material Number (UPN): M00553690;59 unitsClass IOngoing11 Feb 2026
Z-1161-2026HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;266 unitsClass IOngoing11 Feb 2026
Z-1162-2026*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;150 unitsClass IOngoing11 Feb 2026
Z-1163-2026*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 20mm x 10mm; Material Number (UPN): M00553560;588 unitsClass IOngoing11 Feb 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class I
13%
Product records in this event
6Median of the firm: 2
Events of the firm on the register
162

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 98237. https://recallregister.com/fda/i/98237/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France,…
What quantity of product is in FDA event 98237?
The event has 6 product records. The first record gives this quantity: 557 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?
13% of the 945 device events that started in 2025 are in Class I.