Recall Register

Device recall: ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos.…

Class I FDA enforcement event 97318. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Boston Scientific Corporation started this device recall on 24 Jul 2025. The firm address in the FDA record is in Saint Paul, MN. FDA put the recall in Class I. FDA event 97318 has 3 product records. The first FDA enforcement report date is 27 Aug 2025. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution. US nationwide. International distribution worldwide.

Recalling firm

Boston Scientific Corporation, Saint Paul, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 97318: product records
Recall numberProductQuantityClassStatusReport date
Z-2325-2025ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187247,702 US; 344,670 OUSClass IOngoing27 Aug 2025
Z-2326-2025ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296247,702 US; 344,670 OUSClass IOngoing27 Aug 2025
Z-2327-2025RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696247,702 US; 344,670 OUSClass IOngoing27 Aug 2025

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class I
13%
Product records in this event
3Median of the firm: 2
Events of the firm on the register
162

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 97318. https://recallregister.com/fda/i/97318/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution. US nationwide. International distribution worldwide.
What quantity of product is in FDA event 97318?
The event has 3 product records. The first record gives this quantity: 247,702 US; 344,670 OUS
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?
13% of the 945 device events that started in 2025 are in Class I.