Recall Register

Device recall: Baxter EXACTAMIX Inlet, Non-Vented, High Volume Inlet, REF H938173, for use in a pharmaceutical compounding device

Class I FDA enforcement event 95192. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing,…

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Baxter Healthcare Corporation started this device recall on 20 Aug 2024. The firm address in the FDA record is in Deerfield, IL. FDA put the recall in Class I. FDA event 95192 has 4 product records. The first FDA enforcement report date is 25 Sep 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution.

Recalling firm

Baxter Healthcare Corporation, Deerfield, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 95192: product records
Recall numberProductQuantityClassStatusReport date
Z-3082-2024Baxter EXACTAMIX Inlet, Non-Vented, High Volume Inlet, REF H938173, for use in a pharmaceutical compounding device3,580,283 unitsClass IOngoing25 Sep 2024
Z-3083-2024Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pharmaceutical compounding device1,548,782 unitsClass IOngoing25 Sep 2024
Z-3084-2024Baxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a pharmaceutical compounding device5,828,540 unitsClass IOngoing25 Sep 2024
Z-3085-2024Baxter EXACTAMIX Inlet, Vented, Syringe Inlet, REF H938176, for use in a pharmaceutical compounding device2,140,046 unitsClass IOngoing25 Sep 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class I
11%
Product records in this event
4Median of the firm: 1
Events of the firm on the register
193

Other events of the firm

Other FDA enforcement events of Baxter Healthcare Corporation
InitiatedProduct (first record)TypeClassRecords
17 Aug 20260.9% Sodium Chloride Injection USP 500 mL, VIAFLEX Plastic Containers, with 24 units per case. Baxter Healthcare CorporaDrugClass I1
17 Aug 2026Dextrose Injection, USP, 70 %, 2000 mL bags, Rx Only, Baxter Healthcare Corporation Deerfield, IL, 60016, Made in USA, NDrugClass I1
11 Aug 2026Baxter Novum IQ Syringe Pump, Model Number 40800BAXUSDeviceClass II2
6 Aug 2026Vancomycin Injection, USP in 5% Dextrose, 1 g per 200mL (5 mg/mL), in GALAXY Single-Dose Container, Rx only, Sterile NonDrugClass II17
24 Jul 2026Cefazolin in Dextrose, Injection, USP, 2g / 100mL (20mg / mL) Single-Dose Infusion Bag in 100mL GALAXY Plastic ContainerDrugClass I1
30 Jun 2026Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration SetsDeviceClass I1
21 May 2026Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung ExpaDeviceClass I4
22 Apr 2026Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: TubingDeviceClass II1
21 Apr 2026Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper BDeviceClass II1
18 Feb 2026Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software DeviceClass II1

Baxter Healthcare Corporation: all 193 events

Official channels

Cite this page

Recall Register. FDA enforcement event 95192. https://recallregister.com/fda/i/95192/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution.
What quantity of product is in FDA event 95192?
The event has 4 product records. The first record gives this quantity: 3,580,283 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
11% of the 970 device events that started in 2024 are in Class I.