Device recall: Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product…

Class I FDA enforcement event 94701. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Potential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Baxter Healthcare Corporation started this device recall on 29 May 2024. The firm address in the FDA record is in Deerfield, IL. FDA put the recall in Class I. FDA event 94701 has 1 product record. The first FDA enforcement report date is 3 Jul 2024. The record gives the status Ongoing. 371 of the 972 device events that started in 2024 have more product records. The median device event of 2024 has 1 product record. The openFDA file has 972 device events that started in 2024 and 917 that started in 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution.

Recalling firm

Baxter Healthcare Corporation, Deerfield, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 94701: product records
Recall numberProductQuantityClassStatusReport date
Z-2140-2024Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system includes the Life2000 ventilator which has its individual label.2,510 unitsClass IOngoing3 Jul 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2024
972
Share of these events in Class I
11%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
196

Other events of the firm

Other FDA enforcement events of Baxter Healthcare Corporation
InitiatedProduct (first record)TypeClassRecords
14 Sep 2026Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product DescriptioDeviceClass II1
18 Aug 2026PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intakeDeviceClass II3
17 Aug 20260.9% Sodium Chloride Injection USP 500 mL, VIAFLEX Plastic Containers, with 24 units per case. Baxter Healthcare CorporaDrugClass I1
17 Aug 2026Dextrose Injection, USP, 70 %, 2000 mL bags, Rx Only, Baxter Healthcare Corporation Deerfield, IL, 60016, Made in USA, NDrugClass I1
11 Aug 2026Baxter Novum IQ Syringe Pump, Model Number 40800BAXUSDeviceClass II2
6 Aug 2026Vancomycin Injection, USP in 5% Dextrose, 1 g per 200mL (5 mg/mL), in GALAXY Single-Dose Container, Rx only, Sterile NonDrugClass II17
3 Aug 2026Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product DesDeviceClass II1
24 Jul 2026Cefazolin in Dextrose, Injection, USP, 2g / 100mL (20mg / mL) Single-Dose Infusion Bag in 100mL GALAXY Plastic ContainerDrugClass I1
30 Jun 2026Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration SetsDeviceClass I1
21 May 2026Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung ExpaDeviceClass I4

Baxter Healthcare Corporation: all 196 events

Official channels

Cite this page

Recall Register. FDA enforcement event 94701. https://recallregister.com/fda/i/94701/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution.
What quantity of product is in FDA event 94701?
The FDA record gives this quantity: 2,510 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
11% of the 972 device events that started in 2024 are in Class I.