Device recall: Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description…
Class I FDA enforcement event 94690. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
ICU Medical Inc started this device recall on 7 May 2024. The firm address in the FDA record is in Lake Forest, IL. FDA put the recall in Class I. FDA event 94690 has 3 product records. The first FDA enforcement report date is 21 Aug 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa,…
Recalling firm
ICU Medical Inc, Lake Forest, IL.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2447-2024 | Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description is PLUM 360 Infuser. OUS units marketed with the following descriptions: Bomba de infusin Plum 360 compatible con el software ICU Medical MedNet, Plum 360 Infusion System, PLUM 360, PLUM 360"… | 11,961 units | Class I | Ongoing | 21 Aug 2024 |
| Z-2448-2024 | Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 11973. OUS units marketed with the following descriptions: Plum A+ Mednet TM Firmware New 13.6 Span, Plum A+3, PLUM A+3 INFUSION… | 14,744 units | Class I | Ongoing | 21 Aug 2024 |
| Z-2449-2024 | Plum 360/A+ spare batteries. Item Number: SUB0000864. | 130,826 units | Class I | Ongoing | 21 Aug 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class I
- 11%
- Product records in this event
- 3Median of the firm: 2
- Events of the firm on the register
- 75
Other events of the firm
Official channels
- FDA enforcement report Z-2447-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 94690. https://recallregister.com/fda/i/94690/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa,…
- What quantity of product is in FDA event 94690?
- The event has 3 product records. The first record gives this quantity:
11,961 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated