Recall Register

Device recall: Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580…

Class I FDA enforcement event 93327. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

MEDLINE INDUSTRIES, LP - Northfield started this device recall on 16 Oct 2023. The firm address in the FDA record is in Northfield, IL. FDA put the recall in Class I. FDA event 93327 has 3 product records. The first FDA enforcement report date is 3 Jan 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.

Recalling firm

MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 93327: product records
Recall numberProductQuantityClassStatusReport date
Z-0561-2024Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A,…620,735 kitsClass IOngoing3 Jan 2024
Z-0562-2024(1) Nursing Skills kit, SKU #EDUC05038B, Case UPC 40195327019878, Non-sterile, For Training Purposes Only, Not For Human Use.172 kitsClass IOngoing3 Jan 2024
Z-0563-2024(1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile, For Training Purposes Only, Not For Human Use.; and (2) Nursing Tote Kit, SKU #EDUC5056A, Case UPC 40195327153695, Non-sterile, For Training Purposes Only, Not For Human Use.550 kitsClass IOngoing3 Jan 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class I
10%
Product records in this event
3Median of the firm: 2
Events of the firm on the register
120

Other events of the firm

Other FDA enforcement events of MEDLINE INDUSTRIES, LP - Northfield
InitiatedProduct (first record)TypeClassRecords
25 Jul 2025Medline medical convenience kits labeled as: 1. ADULT INTUBATION TRAY, Kit Number ACC010369B; 2. ADULT ICU INTUBATION TRDeviceClass I2
23 Jul 2025Medline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004A LAVH MINOR DeviceClass II1
14 Jul 2025Medline ReNewal Reprocessed St. Jude Medical Response and Supreme Diagnostic Electrophysiology Catheter Models 401150RH,DeviceClass I1
8 Jul 2025The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utiliDeviceClass II1
7 Jul 2025Medline Alcohol Prep Pads, 70% Isopropyl Alcohol, 100 Sterile 2-Ply Pads, Single Use Only, Large, www.medline.com, ManufDrugClass II6
16 Jun 2025SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301DeviceClass II1
12 Jun 2025Medline Convenience Kits, containing BD ChloraPrep Clear - 1mL Applicator, labeled as: LABOR TRIAGE KIT, REF DYKM1564DeviceClass II4
23 May 2025Medline convenience kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN. labeled as: 1) CC DRAWER DeviceClass II3
15 May 2025Medline UMBILICAL TRAY W/3.5&5FR CATH, REF UVT1250DeviceClass II1
22 Apr 2025Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS98DeviceClass I3

MEDLINE INDUSTRIES, LP - Northfield: all 120 events

Official channels

Cite this page

Recall Register. FDA enforcement event 93327. https://recallregister.com/fda/i/93327/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
What quantity of product is in FDA event 93327?
The event has 3 product records. The first record gives this quantity: 620,735 kits
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
10% of the 917 device events that started in 2023 are in Class I.