Recall Register

Device recall: BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE…

Class I FDA enforcement event 93052. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Olympus Corporation of the Americas started this device recall on 12 Oct 2023. The firm address in the FDA record is in Center Valley, PA. FDA put the recall in Class I. FDA event 93052 has 6 product records. The first FDA enforcement report date is 15 Nov 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution

Recalling firm

Olympus Corporation of the Americas, Center Valley, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 93052: product records
Recall numberProductQuantityClassStatusReport date
Z-0189-2024BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT16015942 devicesClass IOngoing15 Nov 2023
Z-0190-2024EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number Q180-AC1154 devicesClass IOngoing15 Nov 2023
Z-0191-2024EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T180, BF-1TQ180, BF-P18013950 devicesClass IOngoing15 Nov 2023
Z-0192-2024EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q17032211 devicesClass IOngoing15 Nov 2023
Z-0193-2024EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190719 devicesClass IOngoing15 Nov 2023
Z-0194-2024EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P1903948 devicesClass IOngoing15 Nov 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class I
10%
Product records in this event
6Median of the firm: 1
Events of the firm on the register
141

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 93052. https://recallregister.com/fda/i/93052/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution
What quantity of product is in FDA event 93052?
The event has 6 product records. The first record gives this quantity: 15942 devices
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
10% of the 917 device events that started in 2023 are in Class I.