Recall Register

Device recall: ADVANTIO DR EL (pacemaker), Model Numbers: a) J064 b) J067 c) K064 (added 7/15/2021) d) K084 (added 7/15/2021) e) K087…

Class I FDA enforcement event 88107. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to…

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Boston Scientific Corporation started this device recall on 3 Jun 2021. The firm address in the FDA record is in Saint Paul, MN. FDA put the recall in Class I. FDA event 88107 has 6 product records. The first FDA enforcement report date is 21 Jul 2021. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution.

Recalling firm

Boston Scientific Corporation, Saint Paul, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 88107: product records
Recall numberProductQuantityClassStatusReport date
Z-2015-2021ADVANTIO DR EL (pacemaker), Model Numbers: a) J064 b) J067 c) K064 (added 7/15/2021) d) K084 (added 7/15/2021) e) K087 (added 7/15/2021)15,717 unitsClass IOngoing21 Jul 2021
Z-2016-2021INGENIO DR EL (pacemakers), Model Numbers: a) J174 b) J177 c) K174 (added 7/15/2021) d) K184 (added 7/15/2021) e) K187 (added 7/15/2021)40,224 unitsClass IOngoing21 Jul 2021
Z-2017-2021VITALIO DR EL (pacemaker), Model Numbers: a) J274 b) J277 c) K274 (added 7/15/2021) d) K277 (added 7/15/2021) e) K284 (added 7/15/2021) f) K287 (added 7/15/2021)16,525 unitsClass IOngoing21 Jul 2021
Z-2018-2021INVIVE CRT-P, Model Numbers: a) V172 b) V173 c) V182 d) V183 e) W172 f) W17313332 unitsClass IOngoing21 Jul 2021
Z-2019-2021INTUA CRT-P, Model Numbers: a) V272 b) V273 c) W2731707 unitsClass IOngoing21 Jul 2021
Z-2020-2021INLIVEN CRT-P, Model Numbers: a) V284 b) V285 c) W274 d) W275 (added 7/15/2021)3634 unitsClass IOngoing21 Jul 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class I
8%
Product records in this event
6Median of the firm: 2
Events of the firm on the register
162

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 88107. https://recallregister.com/fda/i/88107/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution.
What quantity of product is in FDA event 88107?
The event has 6 product records. The first record gives this quantity: 15,717 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
8% of the 900 device events that started in 2021 are in Class I.