Recall Register

Device recall: Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent

Class I FDA enforcement event 87996. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

In rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5 when certain amplification and detection platforms (thermocyclers) are used. These specimens would be falsely interpreted as negative if target amplification occurs prior to a cycle threshold (Ct) value of 5. There have been no formal complaints of False…

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Quidel Corporation started this device recall on 27 Apr 2021. The firm address in the FDA record is in Athens, OH. FDA put the recall in Class I. FDA event 87996 has 1 product record. The first FDA enforcement report date is 7 Jul 2021. The record gives the status Terminated, with the termination date 8 Sep 2022.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution in the states of AL, CA, CT, DC, FL, ID, IN, MA, MI, MO, NM, NY, OH, SC, TN, TX, WA and the countries of Canada, Saudi Arabia, United Kingdom.

Recalling firm

Quidel Corporation, Athens, OH.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 87996: product records
Recall numberProductQuantityClassStatusReport date
Z-1928-2021Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent29,787 kits (96 results/kit)Class ITerminated7 Jul 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class I
8%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
13

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87996. https://recallregister.com/fda/i/87996/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution in the states of AL, CA, CT, DC, FL, ID, IN, MA, MI, MO, NM, NY, OH, SC, TN, TX, WA and the countries of Canada, Saudi Arabia, United Kingdom.
What quantity of product is in FDA event 87996?
The FDA record gives this quantity: 29,787 kits (96 results/kit)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
8% of the 900 device events that started in 2021 are in Class I.