Recall Register

Device recall: Alaris System PC Unit Model 8000 modular infusion pump and monitoring system

Class I FDA enforcement event 85915. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

CareFusion 303, Inc. started this device recall on 30 Jun 2020. The firm address in the FDA record is in San Diego, CA. FDA put the recall in Class I. FDA event 85915 has 2 product records. The first FDA enforcement report date is 12 Aug 2020. The record gives the status Terminated, with the termination date 22 Apr 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT,…

Recalling firm

CareFusion 303, Inc., San Diego, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 85915: product records
Recall numberProductQuantityClassStatusReport date
Z-2740-2020Alaris System PC Unit Model 8000 modular infusion pump and monitoring system37141Class ITerminated12 Aug 2020
Z-2741-2020Alaris System PC Unit Model 8015. modular infusion pump and monitoring system824277Class ITerminated12 Aug 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class I
4%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
99

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 85915. https://recallregister.com/fda/i/85915/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT,…
What quantity of product is in FDA event 85915?
The event has 2 product records. The first record gives this quantity: 37141
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
4% of the 916 device events that started in 2020 are in Class I.