Recall Register

Device recall: Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), 0998-00-0800-53 For cardiovascular use.

Class I FDA enforcement event 82897. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Datascope Corp. started this device recall on 17 Jun 2019. The firm address in the FDA record is in Mahwah, NJ. FDA put the recall in Class I. FDA event 82897 has 6 product records. The first FDA enforcement report date is 24 Jul 2019. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution. US Nationwide, AFGHANISTAN , HUNGARY , PANAMA , ALGERIA , ICELAND , PARAGUAY , ARGENTINA , INDIA , PERU , AUSTRALIA , INDONESIA , PHILIPPINES , AUSTRIA , IRAN , POLAND ,…

Recalling firm

Datascope Corp., Mahwah, NJ.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 82897: product records
Recall numberProductQuantityClassStatusReport date
Z-1939-2019Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), 0998-00-0800-53 For cardiovascular use.6,197Class IOngoing24 Jul 2019
Z-1940-2019Cardiosave Rescue IABP, 0998-00-0800-83 Intra-Aortic Balloon Pump For cardiovascular use.6,197Class IOngoing24 Jul 2019
Z-1941-2019CS100 Intra-Aortic Balloon Pump, English, 110V; 0998-00-3013-53 For cardiovascular use.16,656Class IOngoing24 Jul 2019
Z-1942-2019CS100 Intra-Aortic Balloon Pump, English, 220V; 0998-00-3013-55 For cardiovascular use.16,656Class IOngoing24 Jul 2019
Z-1943-2019CS300 Intra-Aortic Balloon Pump, English, 110V; 0998-00-3023-53 For cardiovascular use.16,656Class IOngoing24 Jul 2019
Z-1944-2019CS300 Intra-Aortic Balloon Pump, English, 220V; 0998-00-3023-55 For cardiovascular use.16,656Class IOngoing24 Jul 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class I
6%
Product records in this event
6Median of the firm: 2
Events of the firm on the register
30

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 82897. https://recallregister.com/fda/i/82897/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution. US Nationwide, AFGHANISTAN , HUNGARY , PANAMA , ALGERIA , ICELAND , PARAGUAY , ARGENTINA , INDIA , PERU , AUSTRALIA , INDONESIA , PHILIPPINES , AUSTRIA , IRAN , POLAND ,…
What quantity of product is in FDA event 82897?
The event has 6 product records. The first record gives this quantity: 6,197
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
6% of the 968 device events that started in 2019 are in Class I.