Recall Register

Food, drugs, devices › GE Healthcare, LLC

Device recall: Giraffe Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are un

FDA enforcement event 82485 · Class I · initiated 15 Mar 2019 · status Terminated · terminated 21 May 2024

DeviceVoluntary: Firm initiated2 product recordsWaukesha, WI

GE Healthcare, LLC (Waukesha, WI) initiated this device recall on 15 Mar 2019; FDA lists 2 product records under event 82485, classified Class I.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 6% of the 968 device enforcement events initiated in 2019 carry the same class.

Status: Terminated, terminated 21 May 2024. Initiation: voluntary: firm initiated. GE Healthcare, LLC has 225 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

GE Healthcare, LLC, Waukesha, WI.

Reason for recall

"The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle."

Distribution

"Worldwide Distribution: US (Nationwide distribution) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1846-2019Giraffe Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open…9,094 devicesClass ITerminated17 Jul 2019
Z-1847-2019Panda Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.28,094 unitsClass ITerminated17 Jul 2019

Context

6%of 2019 device events are Class I968 events that year
2product records in this eventfirm median: 1
225events on this register for the firm

Other enforcement events for GE Healthcare, LLC

InitiatedProduct (first record)TypeClassRecords
27 Feb 2026SIGNA Premier systemsDeviceClass II1
15 Jul 2022(1) GE Centricity Universal Viewer 6.0, Model 2088026-9XX; (2) GE Centricity Universal Viewer 7.0, Model 5826659-0XX; (3DeviceClass II1
24 Jun 2022SIGNA Premier magnetic resonance scanner, model 5748519.DeviceClass II1
14 Jun 2022Centricity PACS Software Version 7.0 SP0.0.4.7DeviceClass II1
8 Jun 2022GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data withinDeviceClass II1
2 Jun 2022Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High CapDeviceClass II1
1 Jun 2022GE Healthcare Aisys CS2 Anesthesia System, Part number 1011-9050-000; and Avance CS2 and Avance CS2 Pro Anesthesia SysteDeviceClass I1
27 May 2022GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, PartDeviceClass II1
10 May 2022CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and DeviceClass II1
28 Apr 2022Centricity Universal Viewer Zero Footprint ClientDeviceClass II5

All 225 events for GE Healthcare, LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle."

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution: US (Nationwide distribution) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 21 May 2024. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.