Recall Register

Device recall: e500 Automatic Transport Ventilator, Model Number 01EVE500

Class I FDA enforcement event 82178. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

O-Two Medical Technologies, Inc. started this device recall on 12 Feb 2019. The firm address in the FDA record is in Brampton, Ontario. FDA put the recall in Class I. FDA event 82178 has 3 product records. The first FDA enforcement report date is 10 Apr 2019. The record gives the status Terminated, with the termination date 29 Jun 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.

Recalling firm

O-Two Medical Technologies, Inc., Brampton, Ontario, Canada.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 82178: product records
Recall numberProductQuantityClassStatusReport date
Z-0976-2019e500 Automatic Transport Ventilator, Model Number 01EVE500136Class ITerminated10 Apr 2019
Z-0977-2019e600 Automatic Transport Ventilator, Model Number 01EVE600119Class ITerminated10 Apr 2019
Z-0978-2019e700 Automatic Transport Ventilator, Model Number 01EVE700632Class ITerminated10 Apr 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class I
6%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
4

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 82178. https://recallregister.com/fda/i/82178/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
What quantity of product is in FDA event 82178?
The event has 3 product records. The first record gives this quantity: 136
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
6% of the 968 device events that started in 2019 are in Class I.