Recall Register

Device recall: arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424. For the in vitro…

Class I FDA enforcement event 72860. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Arkray Factory USA, Inc. started this device recall on 18 Dec 2015. The firm address in the FDA record is in Minneapolis, MN. FDA put the recall in Class I. FDA event 72860 has 2 product records. The first FDA enforcement report date is 27 Jan 2016. The record gives the status Terminated, with the termination date 4 Jan 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.

Recalling firm

Arkray Factory USA, Inc., Minneapolis, MN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 72860: product records
Recall numberProductQuantityClassStatusReport date
Z-0630-2016arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424. For the in vitro quantitative determination of glucose in human plasma from whole blood samples.88 boxes (25 foil packaged test strips per box)Class ITerminated27 Jan 2016
Z-0631-2016arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.11 boxes (25 foil packaged test strips per box)Class ITerminated27 Jan 2016

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class I
4%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
2

Other events of the firm

Other FDA enforcement events of Arkray Factory USA, Inc.
InitiatedProduct (first record)TypeClassRecords
3 May 2019AUTION HYBRID AU-4050 Event 82737DeviceClass II1

Official channels

Cite this page

Recall Register. FDA enforcement event 72860. https://recallregister.com/fda/i/72860/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
What quantity of product is in FDA event 72860?
The event has 2 product records. The first record gives this quantity: 88 boxes (25 foil packaged test strips per box)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?
4% of the 1,153 device events that started in 2015 are in Class I.