Recall Register

Device recall: HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is…

Class I FDA enforcement event 72319. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Hamilton Medical, Inc. started this device recall on 22 Apr 2014. The firm address in the FDA record is in Reno, NV. FDA put the recall in Class I. FDA event 72319 has 1 product record. The first FDA enforcement report date is 11 Nov 2015. The record gives the status Terminated, with the termination date 24 Aug 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide Distribution

Recalling firm

Hamilton Medical, Inc., Reno, NV.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 72319: product records
Recall numberProductQuantityClassStatusReport date
Z-0151-2016HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients1128Class ITerminated11 Nov 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class I
5%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
10

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 72319. https://recallregister.com/fda/i/72319/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide Distribution
What quantity of product is in FDA event 72319?
The FDA record gives this quantity: 1128
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
5% of the 1,176 device events that started in 2014 are in Class I.