Recall Register

Device recall: EnVeo R Loading System, Model No. LS-EnVeoR-23, LS-EnVeoR-2629, LS-EnVeoR-23-C, LS-EnVeoR-2629-C. Cardiovascular valve…

Class I FDA enforcement event 71719. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Medtronic is recalling the EnVeo R Loading Systems because of the presence of particulate.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Medtronic Cardiovascular Surgery-the Heart Valve Division started this device recall on 15 Jul 2015. The firm address in the FDA record is in Santa Ana, CA. FDA put the recall in Class I. FDA event 71719 has 1 product record. The first FDA enforcement report date is 12 Aug 2015. The record gives the status Terminated, with the termination date 20 May 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US Nationwide, Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,…

Recalling firm

Medtronic Cardiovascular Surgery-the Heart Valve Division, Santa Ana, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 71719: product records
Recall numberProductQuantityClassStatusReport date
Z-2264-2015EnVeo R Loading System, Model No. LS-EnVeoR-23, LS-EnVeoR-2629, LS-EnVeoR-23-C, LS-EnVeoR-2629-C. Cardiovascular valve delivery system for use with the Medtronic CoreValve Evolut R system.6,912 units total (540 units in US)Class ITerminated12 Aug 2015

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2015
1,153
Share of these events in Class I
4%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
4

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 71719. https://recallregister.com/fda/i/71719/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US Nationwide, Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,…
What quantity of product is in FDA event 71719?
The FDA record gives this quantity: 6,912 units total (540 units in US)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2015 have the same class?
4% of the 1,153 device events that started in 2015 are in Class I.