Device recall: ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable…

Class I FDA enforcement event 69875. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Heart Sync, Inc. started this device recall on 11 Nov 2014. The firm address in the FDA record is in Ann Arbor, MI. FDA put the recall in Class I. FDA event 69875 has 5 product records. The first FDA enforcement report date is 31 Dec 2014. The record gives the status Terminated, with the termination date 24 Nov 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and…

Recalling firm

Heart Sync, Inc., Ann Arbor, MI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 69875: product records
Recall numberProductQuantityClassStatusReport date
Z-0583-2015ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.sold in boxes of ten electrodes: 4663.8 boxes; 46638 units totalClass ITerminated31 Dec 2014
Z-0584-2015ADULT/CHILD Radiotransparent Electrode, Part number C100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.sold in boxes of ten electrodes: 116 boxes; 1160 units totalClass ITerminated31 Dec 2014
Z-0585-2015ADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.sold in boxes of ten electrodes: 1791.3 boxes; 17913 units totalClass ITerminated31 Dec 2014
Z-0586-2015ADULT Radiotranslucent Electrode, Part number T100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.sold in boxes of ten electrodes: 4692.9 boxes; 46929 units totalClass ITerminated31 Dec 2014
Z-0587-2015ADULT/CHILD Radiotranslucent Electrode, Part number T100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.sold in boxes of ten electrodes: 209 boxes: 2090 units totalClass ITerminated31 Dec 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class I
5%
Product records in this event
5Median of the firm: 5
Events of the firm on the register
2

Other events of the firm

Other FDA enforcement events of Heart Sync, Inc.
InitiatedProduct (first record)TypeClassRecords
26 Apr 2013Heart Sync C100-Physio Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible pacing/ cardioversioDeviceClass II2

Official channels

Cite this page

Recall Register. FDA enforcement event 69875. https://recallregister.com/fda/i/69875/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and…
What quantity of product is in FDA event 69875?
The event has 5 product records. The first record gives this quantity: sold in boxes of ten electrodes: 4663.8 boxes; 46638 units total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
5% of the 1,176 device events that started in 2014 are in Class I.