Recall Register

Device recall: STAT 2 Flow Controller; 80" (203 cm) Appx 14.8 mL, 20 Drop Primary Set w/2 Clave, STAT 2 Flow Controller, Rotating…

Class I FDA enforcement event 69088. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

ICU is recalling the ConMed Stat 2 flow controller because it may deliver a significantly higher flow rate than was intended.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

ICU Medical, Inc. started this device recall on 19 Aug 2014. The firm address in the FDA record is in San Clemente, CA. FDA put the recall in Class I. FDA event 69088 has 4 product records. The first FDA enforcement report date is 1 Oct 2014. The record gives the status Terminated, with the termination date 18 Nov 2014.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - US (nationwide) in the state of WA, OR and the country of South Africa.

Recalling firm

ICU Medical, Inc., San Clemente, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 69088: product records
Recall numberProductQuantityClassStatusReport date
Z-2619-2014STAT 2 Flow Controller; 80" (203 cm) Appx 14.8 mL, 20 Drop Primary Set w/2 Clave, STAT 2 Flow Controller, Rotating Luer, Item No. 011-C9801. Product Usage: IV Administration sets are intended for the infusion and withdrawal of fluids. The devices are provided sterile.3750 unitsClass ITerminated1 Oct 2014
Z-2620-2014STAT 2 Flow Controller; 80" (203 cm) Appx 14.8 mL, 60 Drop Primary Set w/2 Clave, STAT 2 Flow Controller, Rotating Luer, Item No. 011-C9802. Product Usage: IV Administration sets are intended for the infusion and withdrawal of fluids. The devices are provided sterile.1250 unitsClass ITerminated1 Oct 2014
Z-2621-201483" (211 cm) Appx 10.3 mL, 60 Drop Primary Set w/BCV-Clave, Stat 2 Flow Controller, Clave, Rotating Luer, Item No. B9897. Product Usage: IV Administration sets are intended for the infusion and withdrawal of fluids. The devices are provided sterile.100 unitsClass ITerminated1 Oct 2014
Z-2622-201498" (249 cm) Appx 12.3 ml, 60 Drop Primary Set w/Pre-Slit Port, STAT 2 Flow Controller, Rotating Luer, Item No. Z2648. Product Usage: IV Administration sets are intended for the infusion and withdrawal of fluids. The devices are provided sterile.1150 unitsClass ITerminated1 Oct 2014

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2014
1,176
Share of these events in Class I
5%
Product records in this event
4Median of the firm: 2
Events of the firm on the register
75

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 69088. https://recallregister.com/fda/i/69088/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - US (nationwide) in the state of WA, OR and the country of South Africa.
What quantity of product is in FDA event 69088?
The event has 4 product records. The first record gives this quantity: 3750 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2014 have the same class?
5% of the 1,176 device events that started in 2014 are in Class I.