Device recall: Alaris Pump model 8100 with software version 9.1.18 and software upgrade kits with "Delay Until" Option and…
Class I FDA enforcement event 68066. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
CareFusion is recalling the Alaris Pump model 8100 version 9.1.18 because it may have a software issue that results in situation where the pump module will not properly delay an infusion when the "Delay Until" option or "Multidose" feature is used.
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
CareFusion 303, Inc. started this device recall on 23 Apr 2014. The firm address in the FDA record is in San Diego, CA. FDA put the recall in Class I. FDA event 68066 has 1 product record. The first FDA enforcement report date is 21 May 2014. The record gives the status Terminated, with the termination date 1 Mar 2017.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AU, and KW.
Recalling firm
CareFusion 303, Inc., San Diego, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1567-2014 | Alaris Pump model 8100 with software version 9.1.18 and software upgrade kits with "Delay Until" Option and "Multidose" Feature The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using IV administration sets for continuous… | 57,219 units total (55,756 units in US) | Class I | Terminated | 21 May 2014 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2014
- 1,176
- Share of these events in Class I
- 5%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 99
Other events of the firm
Official channels
- FDA enforcement report Z-1567-2014The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 68066. https://recallregister.com/fda/i/68066/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AU, and KW.
- What quantity of product is in FDA event 68066?
- The FDA record gives this quantity:
57,219 units total (55,756 units in US)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2014 have the same class?
- 5% of the 1,176 device events that started in 2014 are in Class I.