Food, drugs, devices › Hospira Inc.
Device recall: The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supp
Hospira Inc. (Lake Forest, IL) initiated this device recall on 3 Feb 2014; FDA lists 2 product records under event 68047, classified Class I.
FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 5% of the 1,176 device enforcement events initiated in 2014 carry the same class.
Status: Terminated, terminated 7 Mar 2017. Initiation: voluntary: firm initiated. Hospira Inc. has 151 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Hospira Inc., Lake Forest, IL.
Reason for recall
"The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply."
Distribution
"Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0095-2015 | The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC. | 978 power supplies | Class I | Terminated | 5 Nov 2014 |
| Z-0096-2015 | The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC. | US Distribution: 4,709 power supplies; Foreign Distribution: 7,425 power supplies | Class I | Terminated | 5 Nov 2014 |
Context
Other enforcement events for Hospira Inc.
All 151 events for Hospira Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply."
What does Class I mean?
FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 7 Mar 2017. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.