Device recall: GE Healthcare, Aisys, Amingo, Avance, and Avance CS2. Intended for volume or pressure control ventilation.

Class I FDA enforcement event 66381. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

GE Healthcare has recently become aware of a potential safety issue involving the Avance, Avance CS2, and Aisys anesthesia devices. While operating in Pressure Control Ventilation – Volume Guarantee (PCV-VG) mode, the affected products may produce an over delivery tidal volume when given a unique sequence of user inputs and a collapsed bellows (partially filled circuit). Exposure to excessiv

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

GE Healthcare, LLC started this device recall on 16 Sep 2013. The firm address in the FDA record is in Waukesha, WI. FDA put the recall in Class I. FDA event 66381 has 1 product record. The first FDA enforcement report date is 23 Oct 2013. The record gives the status Terminated, with the termination date 1 Oct 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico except DE, ME, NM, RI, and VT, and the countries of UNITED ARAB EMIRATES, JAPAN, KAZAKHSTAN, KENYA, REPUBLIC OF KOREA, KUWAIT,…

Recalling firm

GE Healthcare, LLC, Waukesha, WI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 66381: product records
Recall numberProductQuantityClassStatusReport date
Z-0009-2014GE Healthcare, Aisys, Amingo, Avance, and Avance CS2. Intended for volume or pressure control ventilation.10,263 (3,194 US, 7,069 OUS)Class ITerminated23 Oct 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class I
6%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
225

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 66381. https://recallregister.com/fda/i/66381/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution-USA (nationwide) including DC and Puerto Rico except DE, ME, NM, RI, and VT, and the countries of UNITED ARAB EMIRATES, JAPAN, KAZAKHSTAN, KENYA, REPUBLIC OF KOREA, KUWAIT,…
What quantity of product is in FDA event 66381?
The FDA record gives this quantity: 10,263 (3,194 US, 7,069 OUS)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
6% of the 1,236 device events that started in 2013 are in Class I.