Recall Register

Device recall: Zilver PTX Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug…

Class I FDA enforcement event 64975. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Cook, Inc. started this device recall on 18 Apr 2013. The firm address in the FDA record is in Bloomington, IN. FDA put the recall in Class I. FDA event 64975 has 2 product records. The first FDA enforcement report date is 22 May 2013. The record gives the status Terminated, with the termination date 5 Dec 2013.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia

Recalling firm

Cook, Inc., Bloomington, IN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 64975: product records
Recall numberProductQuantityClassStatusReport date
Z-1274-2013Zilver PTX Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal…1,057 units in USAClass ITerminated22 May 2013
Z-1275-2013Zilver PTX Drug-Eluting Peripheral Stent, - 7 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery).…535 units in USAClass ITerminated22 May 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2013
1,236
Share of these events in Class I
6%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
61

Other events of the firm

Other FDA enforcement events of Cook, Inc.
InitiatedProduct (first record)TypeClassRecords
12 Nov 2021Flexor Check-Flo Introducer Ansel Modification Model Number: KCFW-6.0-18/38-45-RB-ANL0-HC Order Number/GPN: G44154 IntenDeviceClass II2
13 Oct 2021Skinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provide DeviceClass II1
8 Oct 2021Transseptal Needle, Trocar. RPN TSNC-18-71.0, TSNC-19-56.0. GPN G02364, G02365. Product consists of a transseptal needleDeviceClass II2
28 Sep 2021Flexor Check-Flo Introducer, Introducer Set, RPN KCFW-6.0-35-45-RB, GPN G09908 The introducer sheath incorporates a hydrDeviceClass II2
4 Jan 2021Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal SDeviceClass II1
24 Nov 2020Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL0-HC, Order Number G56229 Flexor Introducers and Guiding SheathsDeviceClass I21
22 Oct 2020Fixed Core Wire Guide, RPN TSCF-35-260-3, GPN G00517. The product is a curved wire guide with PTFE coating manufactured DeviceClass I1
31 Aug 2020Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 DeviceClass II2
11 Aug 2020Zenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatment of patients wDeviceClass II1
30 Jul 2020Cook Medical Universa Soft Ureteral Stent Sets-5 FR 26 cm Reference part Number: UFH-526 - Product Usage: Used for tempoDeviceClass II1

Cook, Inc.: all 61 events

Official channels

Cite this page

Recall Register. FDA enforcement event 64975. https://recallregister.com/fda/i/64975/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia
What quantity of product is in FDA event 64975?
The event has 2 product records. The first record gives this quantity: 1,057 units in USA
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2013 have the same class?
6% of the 1,236 device events that started in 2013 are in Class I.