Recall Register

Device recall: GE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator, emergency…

Class I FDA enforcement event 63761. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

GE Healthcare, LLC started this device recall on 15 Oct 2012. The firm address in the FDA record is in Waukesha, WI. FDA put the recall in Class I. FDA event 63761 has 2 product records. The first FDA enforcement report date is 20 Feb 2013. The record gives the status Terminated, with the termination date 18 Jan 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA, MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX, VT, WA, WI and WV; and the…

Recalling firm

GE Healthcare, LLC, Waukesha, WI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 63761: product records
Recall numberProductQuantityClassStatusReport date
Z-0748-2013GE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants.276 units total, 109 of those nationwide.Class ITerminated20 Feb 2013
Z-0749-2013Giraffe and Panda Bag and Mask Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants.276 units total, 109 of those nationwide.Class ITerminated20 Feb 2013

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class I
5%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
225

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 63761. https://recallregister.com/fda/i/63761/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA, MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX, VT, WA, WI and WV; and the…
What quantity of product is in FDA event 63761?
The event has 2 product records. The first record gives this quantity: 276 units total, 109 of those nationwide.
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?
5% of the 823 device events that started in 2012 are in Class I.