Recall Register

Device recall: The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for…

Class I FDA enforcement event 61835. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The Olympic Cool-Cap's Control Module has experienced a frozen screen during use. When this occurs, the on-screen information remains on display, but the system is no longer providing cooling treatment to the infants. The visible clock displayed at the upper right hand corner of the screen stops advancing.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Natus Medical Incorporated started this device recall on 9 May 2012. The firm address in the FDA record is in Seattle, WA. FDA put the recall in Class I. FDA event 61835 has 1 product record. The first FDA enforcement report date is 12 Dec 2012. The record gives the status Terminated, with the termination date 27 Jan 2015.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution: USA (nationwide) including states of: AK, AR, AZ, CA, CO, DE, FL, GA, HI, ID, IL, IN, KY, MA, MI, MN, MS, ND, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, WA, Puerto Rico,…

Recalling firm

Natus Medical Incorporated, Seattle, WA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 61835: product records
Recall numberProductQuantityClassStatusReport date
Z-0447-2013The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for treatment of hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored…324 unitsClass ITerminated12 Dec 2012

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2012
823
Share of these events in Class I
5%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
16

Other events of the firm

Other FDA enforcement events of Natus Medical Incorporated
InitiatedProduct (first record)TypeClassRecords
28 Nov 2023External Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx onlyDeviceClass II2
18 Nov 2022The Natus Cranial Access Kit is designed to be a complete, single use and disposable accessory and instrumentation set tDeviceClass II1
22 May 2017The Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and high-level disinfection of resDeviceClass II1
3 Apr 2017Second generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part Number 010066. TopDeviceClass II1
28 Oct 2016Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatal HyperbiliruDeviceClass II1
16 Sep 2016Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used iDeviceClass II1
25 Aug 2016The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003; KITNC-STLX00DeviceClass II1
22 Apr 2016Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 1DeviceClass II1
11 Apr 2016Natus neoBLUE Phototherapy system, Catalogue Part Number: 010066. Top Assembly Part Numbers: 001103, 001314, 002110, 040DeviceClass II1
8 Jan 2016Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.DeviceClass II1

Natus Medical Incorporated: all 16 events

Official channels

Cite this page

Recall Register. FDA enforcement event 61835. https://recallregister.com/fda/i/61835/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution: USA (nationwide) including states of: AK, AR, AZ, CA, CO, DE, FL, GA, HI, ID, IL, IN, KY, MA, MI, MN, MS, ND, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, WA, Puerto Rico,…
What quantity of product is in FDA event 61835?
The FDA record gives this quantity: 324 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2012 have the same class?