Device recall: Smiths Medical, Medfusion Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers…
Class I FDA enforcement event 61247. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Smiths Medical has identified a software anomaly in the Medfusion 4000 Syringe Infusion Pump that causes a device history log corruption and triggers a Watchdog Fail-Safe” alarm. When the device exhibits this failure mode, visual and audible alarms will sound and the device ceases operation.
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Smiths Medical ASD, Inc. started this device recall on 14 Feb 2012. The firm address in the FDA record is in Saint Paul, MN. FDA put the recall in Class I. FDA event 61247 has 1 product record. The first FDA enforcement report date is 8 Aug 2012. The record gives the status Terminated, with the termination date 14 Nov 2012.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution -- USA, including the state of CA, and country of Canada.
Recalling firm
Smiths Medical ASD, Inc., Saint Paul, MN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2107-2012 | Smiths Medical, Medfusion Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers 4000-0100-50, 4000-0100-95, 4000-0101-50, 4000-0101-51, and 4000-0101-78. Designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. | 607 pumps (257 USA, 350 Canada) | Class I | Terminated | 8 Aug 2012 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2012
- 823
- Share of these events in Class I
- 5%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 108
Other events of the firm
Official channels
- FDA enforcement report Z-2107-2012The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 61247. https://recallregister.com/fda/i/61247/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution -- USA, including the state of CA, and country of Canada.
- What quantity of product is in FDA event 61247?
- The FDA record gives this quantity:
607 pumps (257 USA, 350 Canada)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2012 have the same class?