Device recall: ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the

FDA enforcement event 99596 · Class II · initiated 6 Aug 2026 · status Ongoing

DeviceVoluntary: Firm initiated2 product recordsChaska, MN

FDA enforcement event 99596 is a device recall by IMRIS Imaging Inc of Chaska, MN, initiated 6 Aug 2026 and classified Class II, with 2 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 572 device enforcement events initiated in 2026 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. IMRIS Imaging Inc has 2 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

IMRIS Imaging Inc, Chaska, MN.

Reason for recall

"The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure."

Distribution

"Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden."

Product records

Recall numberProductQuantityClassStatusReport date
Z-3033-2026ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-0006 unitsClass IIOngoing2 Sep 2026
Z-3034-2026ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-60013 unitsClass IIOngoing2 Sep 2026

Context

87%of 2026 device events are Class II572 events that year
2product records in this eventfirm median: 2
2events on this register for the firm

Other enforcement events for IMRIS Imaging Inc

InitiatedProduct (first record)TypeClassRecords
8 Dec 2025Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000, 119630-000, 1 (98132)DeviceClass II1

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.