Device recall: Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
FDA enforcement event 98052 is a device recall by Philips North America of Cambridge, MA, initiated 27 Oct 2025 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 85% of the 945 device enforcement events initiated in 2025 carry the same class.
Status: Ongoing. Initiation: fda mandated. Philips North America has 30 enforcement events on this register, with a median of 2 product records per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Philips North America, Cambridge, MA.
Reason for recall
"Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications."
Distribution
"U.S."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0597-2026 | Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM | 110 | Class II | Ongoing | 17 Dec 2025 |
Context
Other enforcement events for Philips North America
All 30 events for Philips North America
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "U.S."
Is the recall still ongoing?