Recall Register

Food, drugs, devices › Vantive US Healthcare LLC

Device recall: Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8

FDA enforcement event 97527 · Class II · initiated 28 Aug 2025 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordDeerfield, IL

FDA enforcement event 97527 is a device recall by Vantive US Healthcare LLC of Deerfield, IL, initiated 28 Aug 2025 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 85% of the 945 device enforcement events initiated in 2025 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Vantive US Healthcare LLC has 3 enforcement events on this register, with a median of 3 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Vantive US Healthcare LLC, Deerfield, IL.

Reason for recall

"Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the…"

Distribution

"Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0409-2026Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)1Class IIOngoing5 Nov 2025

Context

85%of 2025 device events are Class II945 events that year
1product record in this eventfirm median: 3
3events on this register for the firm

Other enforcement events for Vantive US Healthcare LLC

InitiatedProduct (first record)TypeClassRecords
6 Jan 2026PRISMAFLEX ST100 SET US, Product Code 107636US; Dialyzer, High Permeability With Or Without Sealed Dialysate SystemDeviceClass II8
28 Jul 2025PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701DeviceClass II3

All 3 events for Vantive US Healthcare LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the…"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.