Device recall: Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
GE Medical Systems, LLC (Waukesha, WI) initiated this device recall on 28 Jan 2025; FDA lists 4 product records under event 96208, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 85% of the 945 device enforcement events initiated in 2025 carry the same class.
Status: Ongoing. Initiation: fda mandated. GE Medical Systems, LLC has 42 enforcement events on this register, with a median of 3 product records per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
GE Medical Systems, LLC, Waukesha, WI.
Reason for recall
"The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached."
Distribution
"Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1043-2025 | Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system. | 3149 (1697 US; 1452 OUS) | Class II | Ongoing | 12 Feb 2025 |
| Z-1044-2025 | Discovery XR656HD. X-Ray imaging system. | 3149 (1697 US; 1452 OUS) | Class II | Ongoing | 12 Feb 2025 |
| Z-1045-2025 | Discovery XR656 HD. X-Ray imaging system. | 3149 (1697 US; 1452 OUS) | Class II | Ongoing | 12 Feb 2025 |
| Z-1046-2025 | Optima XR646 HD. X-Ray imaging system. | 3149 (1697 US; 1452 OUS) | Class II | Ongoing | 12 Feb 2025 |
Context
Other enforcement events for GE Medical Systems, LLC
All 42 events for GE Medical Systems, LLC
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican…"
Is the recall still ongoing?
FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.