Recall Register

Device recall: PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide…

Class II FDA enforcement event 95173. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Fisher & Paykel Healthcare, Ltd. started this device recall on 7 Aug 2024. The firm address in the FDA record is in Auckland. FDA put the recall in Class II. FDA event 95173 has 1 product record. The first FDA enforcement report date is 2 Oct 2024. The record gives the status Ongoing. 371 of the 972 device events that started in 2024 have more product records. The median device event of 2024 has 1 product record. The openFDA file has 972 device events that started in 2024 and 917 that started in 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided

Recalling firm

Fisher & Paykel Healthcare, Ltd., Auckland, New Zealand.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 95173: product records
Recall numberProductQuantityClassStatusReport date
Z-3262-2024PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.944 unitsClass IIOngoing2 Oct 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
972
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
7

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 95173. https://recallregister.com/fda/ii/95173/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided
What quantity of product is in FDA event 95173?
The FDA record gives this quantity: 944 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?