Recall Register

Food, drugs, devices › Fisher & Paykel Healthcare, Ltd.

Device recall: PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide h

FDA enforcement event 95173 · Class II · initiated 7 Aug 2024 · status Ongoing

DeviceVoluntary: Firm initiated1 product record

Fisher & Paykel Healthcare, Ltd. (Auckland) initiated this device recall on 7 Aug 2024; FDA lists 1 product record under event 95173, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 970 device enforcement events initiated in 2024 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Fisher & Paykel Healthcare, Ltd. has 7 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Fisher & Paykel Healthcare, Ltd., Auckland, New Zealand.

Reason for recall

"Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia."

Distribution

"U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided"

Product records

Recall numberProductQuantityClassStatusReport date
Z-3262-2024PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.944 unitsClass IIOngoing2 Oct 2024

Context

87%of 2024 device events are Class II970 events that year
1product record in this eventfirm median: 1
7events on this register for the firm

Other enforcement events for Fisher & Paykel Healthcare, Ltd.

InitiatedProduct (first record)TypeClassRecords
18 Sep 2025Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723DeviceClass II1
2 Apr 2024Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100USDeviceClass II1
6 Nov 2014Fisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer Fisher and Paykel; Healthcare CosyCot and Kit Replacement DeviceClass II1
15 Apr 2014Infant Nasal CPAP (continuous positive airway pressure) Prong, Nasal. CPAP Prongs (10 pack): Model No. BC3020-10, BC3520DeviceClass I1
22 Apr 2013Fisher & Paykel Healthcare ICON CPAP (continuous positive airway pressure). Models: ICON Auto: ICONAAN, ICONAAN-HT, ICONDeviceClass II1
19 Nov 2012Fisher & Paykel Healthcare reusable breathing circuit. Non-heated breathing circuit, 1.8m or 6 feet in length. Oxygen thDeviceClass I1

All 7 events for Fisher & Paykel Healthcare, Ltd.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.