Device recall: Brand Name: CLEANER BOTTLE Product Name: CLEANER BOTTLE Model/Catalog Number: EG-1000 Software Version: N/A Product…
Class II FDA enforcement event 94293. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Received complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or storage if the use is not releasing the pressure inside the bottles).
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Electro Medical Systems SA started this device recall on 13 Mar 2024. The firm address in the FDA record is in Nyon. FDA put the recall in Class II. FDA event 94293 has 3 product records. The first FDA enforcement report date is 1 May 2024. The record gives the status Ongoing. 166 of the 972 device events that started in 2024 have more product records. The median device event of 2024 has 1 product record. The openFDA file has 972 device events that started in 2024 and 917 that started in 2023.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - U.S. Nationwide in the states of IL and TX. The countries of ALBANIA, ARMENIA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC,…
Recalling firm
Electro Medical Systems SA, Nyon, N/A, Switzerland.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1686-2024 | Brand Name: CLEANER BOTTLE Product Name: CLEANER BOTTLE Model/Catalog Number: EG-1000 Software Version: N/A Product Description: CLEANER bottles used with EMS AIRFLOW PROPHYLAXIS MASTER / EMS AIRFLOW ONE devices to clean the water lines. The bottle is placed on the device during cleaning. Component: YES. AIRFLOW… | 342 | Class II | Ongoing | 1 May 2024 |
| Z-1687-2024 | Brand Name: WATER BOTTLE Product Name: WATER BOTTLE Model/Catalog Number: EG-121 Software Version: N/A Product Description: These bottles are sold empty and are filled by the user with water fort dental treatments. The bottle is pressurized on the device during use. Component: YES. AIRFLOW PROPHYLAXIS MASTER and… | 462 | Class II | Ongoing | 1 May 2024 |
| Z-1688-2024 | Brand Name: PIEZON Bottle Product Name: PIEZON Bottle Model/Catalog Number: EG-111 Software Version: N/A Product Description: These bottles are sold empty and are filled by the user with water - in some cases small quantities of disinfectant solutions - for dental treatments. The bottle is pressurized on the device… | 54 | Class II | Ongoing | 1 May 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 972
- Share of these events in Class II
- 87%
- Product records in this event
- 3
- Events of the firm on the register
- 1
Official channels
- FDA enforcement report Z-1686-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 94293. https://recallregister.com/fda/ii/94293/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - U.S. Nationwide in the states of IL and TX. The countries of ALBANIA, ARMENIA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC,…
- What quantity of product is in FDA event 94293?
- The event has 3 product records. The first record gives this quantity:
342
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 972 device events that started in 2024 are in Class II.