Device recall: Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6
Medtronic Inc (Minneapolis, MN) initiated this device recall on 12 Oct 2023; FDA lists 3 product records under event 93258, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Medtronic Inc has 47 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Medtronic Inc, Minneapolis, MN.
Reason for recall
"The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect…"
Distribution
"Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0320-2024 | Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6 | 2767 units | Class II | Ongoing | 29 Nov 2023 |
| Z-0321-2024 | Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130069X6 | 173 units | Class II | Ongoing | 29 Nov 2023 |
| Z-0322-2024 | Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5 | 48 units | Class II | Ongoing | 29 Nov 2023 |
Context
Other enforcement events for Medtronic Inc
All 47 events for Medtronic Inc
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect…"
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom."
Is the recall still ongoing?