Recall Register

Food, drugs, devices › Alphatec Spine, Inc.

Device recall: LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, T

FDA enforcement event 92776 · Class II · initiated 7 Jul 2023 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordCarlsbad, CA

Alphatec Spine, Inc. (Carlsbad, CA) initiated this device recall on 7 Jul 2023; FDA lists 1 product record under event 92776, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Alphatec Spine, Inc. has 17 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Alphatec Spine, Inc., Carlsbad, CA.

Reason for recall

"Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery."

Distribution

"US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI."

Product records

Recall numberProductQuantityClassStatusReport date
Z-2475-2023LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Battalion LLIF,29Class IIOngoing6 Sep 2023

Context

88%of 2023 device events are Class II917 events that year
1product record in this eventfirm median: 1
17events on this register for the firm

Other enforcement events for Alphatec Spine, Inc.

InitiatedProduct (first record)TypeClassRecords
13 May 2026SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix SystemDeviceClass II1
2 Jan 2026ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 266-01-000-10-N UDI-DI code: 00190376476179; GDeviceClass II2
15 Dec 2024Calibrate CCX Interbody System for spinal fusion procedures: Implants: (1) Calibrate CCX-O, 7 x 10 x 25mm 5 degrees, ParDeviceClass II1
26 Jan 2024IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 2DeviceClass II1
14 Sep 2022Disc Preparation instruments contained in the following products: CI-11050-08102515 Longer IdentiTi PO Trial Tip, 11A X DeviceClass II1
26 Apr 2021Atec Insignia Anterior Cervical Plate System, REF 136-0116, Insignia, ACP, 1-Level, 16 mm, Rx Only, Non-Sterile, UDI: (0DeviceClass II29
5 May 2017Arsenal Spinal Fixation System, Set Screw, Part No. 47027 Product Usage: The Arsenal Spinal Fixation System is intended DeviceClass II1
5 Oct 2016Arsenal Spinal Fixation System, Set Screw, Part No. 47127 Product Usage: Usage: The Arsenal Spinal Fixation System is inDeviceClass II1
10 Apr 20154.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm Alphatec Spine Solutions for the Aging Spine Trestle Luxe AnterDeviceClass II1
26 Jun 2014Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization System, aDeviceClass II1

All 17 events for Alphatec Spine, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.