Food, drugs, devices › Smith & Nephew, Inc.
Device recall: Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1
Smith & Nephew, Inc. (Memphis, TN) initiated this device recall on 2 May 2023; FDA lists 4 product records under event 92286, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Smith & Nephew, Inc. has 107 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Smith & Nephew, Inc., Memphis, TN.
Reason for recall
"Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision."
Distribution
"US Nationwide distribution."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1670-2023 | Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); c) REF 1-10003-200 (SIZE 2-Right Medial); d) REF 1-10003-210 (SIZE 2-Left Medial); e) REF 1-10003-300 (SIZE 3- Right Medial); f) REF… | 1994 units | Class II | Ongoing | 7 Jun 2023 |
| Z-1671-2023 | Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6); Unicondylar knee prothesis | 1596 units | Class II | Ongoing | 7 Jun 2023 |
| Z-1672-2023 | Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-10012-200, SIZE 2-LEFT MEDIAL; d) REF 1-10012-250, SIZE 2-RIGHT MEDIAL; e) REF 1-10012-300, SIZE 3-LEFT MEDIAL; f) REF 1-10012-350, SIZE… | 2146 units | Class II | Ongoing | 7 Jun 2023 |
| Z-1673-2023 | Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c) REF 1-10013-111, SIZE 1-LEFT MED, 11MM; d) REF 1-10013-113, SIZE 1-LEFT MED, 13MM; e) REF 1-10013-159, SIZE 1-RIGHT MED, 9MM; f) REF… | 7531 units | Class II | Ongoing | 7 Jun 2023 |
Context
Other enforcement events for Smith & Nephew, Inc.
All 107 events for Smith & Nephew, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US Nationwide distribution."
Is the recall still ongoing?
FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.