Recall Register

Food, drugs, devices › Beckman Coulter, Inc.

Device recall: iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Number

FDA enforcement event 92056 · Class II · initiated 5 Apr 2023 · status Ongoing

DeviceVoluntary: Firm initiated4 product recordsMiami, FL

FDA enforcement event 92056 is a device recall by Beckman Coulter, Inc. of Miami, FL, initiated 5 Apr 2023 and classified Class II, with 4 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Beckman Coulter, Inc. has 216 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Beckman Coulter, Inc., Miami, FL.

Reason for recall

"Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated."

Distribution

"Worldwide Distribution"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1596-2023iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-33251068 unitsClass IIOngoing24 May 2023
Z-1597-2023DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis107 unitsClass IIOngoing24 May 2023
Z-1598-2023DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis355 unitsClass IIOngoing24 May 2023
Z-1599-2023UA WIN 10 Kit, Catalog Number C52900, in-vitro diagnostic device, urine analysis306 unitsClass IIOngoing24 May 2023

Context

88%of 2023 device events are Class II917 events that year
4product records in this eventfirm median: 1
216events on this register for the firm

Other enforcement events for Beckman Coulter, Inc.

InitiatedProduct (first record)TypeClassRecords
15 Jul 2026Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC DeviceClass II2
26 Jun 2026REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-28DeviceClass II2
16 Mar 2026Access Total T4 Calibrator, Catalog No. 33805DeviceClass II1
23 Dec 2025CellMek SPS Sample Preparation System, REF: C44603, with softwareDeviceClass II1
26 Nov 2025Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin SubunitDeviceClass II2
10 Nov 2025COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunctioDeviceClass II1
7 Nov 2025Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibratioDeviceClass II1
1 Oct 2025Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry AnalyzerDeviceClass II2
22 Sep 2025DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use asDeviceClass II1
19 Sep 2025DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i DeviceClass II2

All 216 events for Beckman Coulter, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.