Recall Register

Food, drugs, devices › Boston Scientific Corporation

Device recall: ORISE Gel Submucosal Lifting Agent--Indicated for use in gastrointestinal endoscopic procedures for submucosal lift of p

FDA enforcement event 90986 · Class II · initiated 10 Oct 2022 · status Ongoing

DeviceVoluntary: Firm initiated3 product recordsMarlborough, MA

FDA enforcement event 90986 is a device recall by Boston Scientific Corporation of Marlborough, MA, initiated 10 Oct 2022 and classified Class II, with 3 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 839 device enforcement events initiated in 2022 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Boston Scientific Corporation has 162 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Boston Scientific Corporation, Marlborough, MA.

Reason for recall

"***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention"

Distribution

"Worldwide Distribution: US (Nationwide) and (Foreign) countries of: Algeria Argentina Argentina Australia Austria Bahrain Belgium Brazil Canada Chile Colombia Costa Rica Cyprus Czech Republic…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0224-2023ORISE Gel Submucosal Lifting Agent--Indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device Packaged as follows: UPN: (1)Twin Pack Box 1…402,968 TotalClass IIOngoing16 Nov 2022
Z-0225-2023ORISE Gel Syringe Submucosal Lifting Agent Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device UPN: (1)Twin Pack Box 1 M00519210;…402,968 TotalClass IIOngoing16 Nov 2022
Z-0226-2023ORISE ProKnife Procedure Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device. UPN (1)ORISE ProKnife 1.5 mm Electrode - Kit M00519380;…402,968 TotalClass IIOngoing16 Nov 2022

Context

88%of 2022 device events are Class II839 events that year
3product records in this eventfirm median: 2
162events on this register for the firm

Other enforcement events for Boston Scientific Corporation

InitiatedProduct (first record)TypeClassRecords
6 Aug 2026LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All VerDeviceClass II3
22 Jul 2026Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DeviceClass II1
14 Jul 2026OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device.DeviceClass II2
9 Jul 2026Boston Scientific ENROUTE Transcarotid Neuroprotection System, Material Number SR-200-NPS; Percutaneous catheterDeviceClass I2
30 Jun 2026Rapid Refill Continuous Injection System, UPN M00566001DeviceClass II1
30 Jun 2026Radial Jaw 4 Standard Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 SC 240cm Box 20; DeviceClass II3
19 Jun 2026Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing SysDeviceClass II1
11 Jun 2026FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, MDeviceClass II4
11 Jun 2026FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation CathetDeviceClass II3
11 Jun 2026FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation CathetDeviceClass II1

All 162 events for Boston Scientific Corporation

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution: US (Nationwide) and (Foreign) countries of: Algeria Argentina Argentina Australia Austria Bahrain Belgium Brazil Canada Chile Colombia Costa Rica Cyprus Czech Republic…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.