Recall Register

Food, drugs, devices › Cook Incorporated

Device recall: Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN)RFPC-35-180 ORDER NUMBER (GPN) G07937. For vascular cath

FDA enforcement event 90710 · Class II · initiated 19 Jul 2022 · status Ongoing

DeviceVoluntary: Firm initiated13 product recordsBloomington, IN

Cook Incorporated (Bloomington, IN) initiated this device recall on 19 Jul 2022; FDA lists 13 product records under event 90710, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 839 device enforcement events initiated in 2022 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Cook Incorporated has 16 enforcement events on this register, with a median of 2 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Cook Incorporated, Bloomington, IN.

Reason for recall

"Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised."

Distribution

"Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1706-2022Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN)RFPC-35-180 ORDER NUMBER (GPN) G07937. For vascular catheter positioning and exchange in diagnostic and interventional procedures.496 unitsClass IIOngoing21 Sep 2022
Z-1707-2022Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RFSPC-35-180 ORDER NUMBER (GPN) G09608. For vascular catheter positioning and exchange in diagnostic and interventional procedures.2407 unitsClass IIOngoing21 Sep 2022
Z-1708-2022Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER NUMBER (GPN) G07516. For vascular catheter positioning and exchange in diagnostic and interventional procedures.1450 unitsClass IIOngoing21 Sep 2022
Z-1709-2022Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.1997 unitsClass IIOngoing21 Sep 2022
Z-1710-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPW-35-150 ORDER NUMBER (GPN):G56149. For vascular catheter positioning and exchange in diagnostic and interventional procedures.585 unitsClass IIOngoing21 Sep 2022
Z-1711-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-18-150 ORDER NUMBER (GPN):G56161. For vascular catheter positioning and exchange in diagnostic and interventional procedures.85 unitsClass IIOngoing21 Sep 2022
Z-1712-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-18-80 ORDER NUMBER (GPN): G56162. For vascular catheter positioning and exchange in diagnostic and interventional procedures.35 unitsClass IIOngoing21 Sep 2022
Z-1713-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-150 ORDER NUMBER (GPN): G56173. For vascular catheter positioning and exchange in diagnostic and interventional procedures.5400 unitsClass IIOngoing21 Sep 2022
Z-1714-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-80 ORDER NUMBER (GPN): G56172. For vascular catheter positioning and exchange in diagnostic and interventional procedures.75 unitsClass IIOngoing21 Sep 2022
Z-1715-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-150 ORDER NUMBER (GPN): G56176. For vascular catheter positioning and exchange in diagnostic and interventional procedures.865 unitsClass IIOngoing21 Sep 2022
Z-1716-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-80 ORDER NUMBER (GPN): G56175. For vascular catheter positioning and exchange in diagnostic and interventional procedures.189 unitsClass IIOngoing21 Sep 2022
Z-1717-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-150 ORDER NUMBER (GPN): G56152. For vascular catheter positioning and exchange in diagnostic and interventional procedures.30 unitsClass IIOngoing21 Sep 2022
Z-1718-2022Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-80 ORDER NUMBER (GPN): G56151. For vascular catheter positioning and exchange in diagnostic and interventional procedures.15 unitsClass IIOngoing21 Sep 2022

Context

88%of 2022 device events are Class II839 events that year
13product records in this eventfirm median: 2
16events on this register for the firm

Other enforcement events for Cook Incorporated

InitiatedProduct (first record)TypeClassRecords
2 Apr 2026Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are DeviceClass I3
5 Mar 2026COOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786DeviceClass II18
26 Nov 2025Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9.0-18/38-45-RB-DeviceClass II3
15 May 2025Beacon Tip 5.0 Fr Angiographic catheters are designed to facilitate diagnostic and therapeutic procedures. Catheters areDeviceClass I1
17 Apr 2025Tornado Embolization Microcoil, intended for embolization of selective vessel supply to arteriovenous malformations and DeviceClass II1
17 Apr 2025Check=Flo Performer Introducer, 10.0Fr, Reference Part Number(RPN) REF RCFW-10.0-0-38-30-RB, Order Number (GPN) REF G086DeviceClass II2
14 Oct 2024Approach CTO Micro Wire Guide, RPN CMW-14-190-25G, GPN G50789DeviceClass II1
8 Oct 2024Approach CTO Micro Wire Guide, Guide Wire, Catalog Prefix CMW; Reference Part Numbers: CMW-14-190-12G CMW-14-190-18G CMWDeviceClass II2
16 Sep 2024Check-Flo Performer Introducer Set, Reference Part Number RCF-5.0-35-45-J, Order Number G27242. Performer Introducers arDeviceClass II1
15 May 2024Ultrathane Mac-Loc Locking Loop Multipurpose Drainage Catheter ULT16.0-38-25-P-6S-MCL-HC--Intended for percutaneous draiDeviceClass II2

All 16 events for Cook Incorporated

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.