Device recall: Olympus CYSTONEPHROFIBERSCOPE Model: CYF-5
Class II FDA enforcement event 90195. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Olympus Corporation of the Americas started this device recall on 20 Apr 2022. The firm address in the FDA record is in Center Valley, PA. FDA put the recall in Class II. FDA event 90195 has 3 product records. The first FDA enforcement report date is 22 Jun 2022. The record gives the status Ongoing. 92 of the 839 device events that started in 2022 have more product records. The median device event of 2022 has 1 product record. The openFDA file has 839 device events that started in 2022 and 900 that started in 2021.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
CA FL MA MD MI NJ OH PR WI
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1262-2022 | Olympus CYSTONEPHROFIBERSCOPE Model: CYF-5 | 7 units | Class II | Ongoing | 22 Jun 2022 |
| Z-1263-2022 | Olympus URETERO-RENO FIBERSCOPE Model: URF-P6 | 2 units | Class II | Ongoing | 22 Jun 2022 |
| Z-1264-2022 | Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R | 4 units | Class II | Ongoing | 22 Jun 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class II
- 88%
- Product records in this event
- 3Median of the firm: 1
- Events of the firm on the register
- 145
Other events of the firm
Official channels
- FDA enforcement report Z-1262-2022The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 90195. https://recallregister.com/fda/ii/90195/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
CA FL MA MD MI NJ OH PR WI
- What quantity of product is in FDA event 90195?
- The event has 3 product records. The first record gives this quantity:
7 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?
- 88% of the 839 device events that started in 2022 are in Class II.