Recall Register

Food, drugs, devices › Microbiologics Inc

Device recall: KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 5

FDA enforcement event 89401 · Class II · initiated 18 Feb 2020 · status Terminated · terminated 11 Jul 2022

DeviceVoluntary: Firm initiated5 product recordsSaint Cloud, MN

FDA enforcement event 89401 is a device recall by Microbiologics Inc of Saint Cloud, MN, initiated 18 Feb 2020 and classified Class II, with 5 product records.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 916 device enforcement events initiated in 2020 carry the same class.

Status: Terminated, terminated 11 Jul 2022. Initiation: voluntary: firm initiated. Microbiologics Inc has 43 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Microbiologics Inc, Saint Cloud, MN.

Reason for recall

"Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release."

Distribution

"Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0585-2022KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 512991,975 unitsClass IITerminated16 Feb 2022
Z-0586-2022LYFO-DISK Enterococcus faecalis derived from ATCC 51299 In Vitro Diagnostic Control microorganisms183 unitsClass IITerminated16 Feb 2022
Z-0587-2022QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms10 unitsClass IITerminated16 Feb 2022
Z-0588-2022QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms139 unitsClass IITerminated16 Feb 2022
Z-0589-2022QC Sets and Panels, KWIK-STIK Blood Culture Identification (BCID) Verification Panel (Live Culture). KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms3 unitsClass IITerminated16 Feb 2022

Context

93%of 2020 device events are Class II916 events that year
5product records in this eventfirm median: 1
43events on this register for the firm

Other enforcement events for Microbiologics Inc

InitiatedProduct (first record)TypeClassRecords
2 Dec 2025KWIK-STIK 2-Pack, Campylobacter jejuni subsp. jejuni derived from ATCC 29428, Catalog Number 0325PDeviceClass II3
13 Nov 2025KWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from ATCC 17802, Catalog Number 0818KDeviceClass II1
27 Jun 2025LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)DeviceClass II2
17 Jun 2025KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805LDeviceClass II1
27 May 2025KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC¿ 17960;DeviceClass II1
20 Feb 2025Helix Elite" Inactivated Standard: Inactivated SARS-CoV-2 Whole Virus (Pellet), Multi-analyte, unassayed quality controlDeviceClass II1
20 Feb 2025Helix Elite Inactivated Standard: Inactivated Influenza A/B and Respiratory Syncytial Virus, Multi-analyte, unassayed quDeviceClass II1
20 Jan 2025Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnosticDeviceClass II2
6 Mar 2024KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b, a) REF 0209P and b) REF 0209K, Quality control kit for culturDeviceClass II1
8 Jan 2024Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.DeviceClass II1

All 43 events for Microbiologics Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 11 Jul 2022. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.