Device recall: KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC…
Class II FDA enforcement event 89401. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Microbiologics Inc started this device recall on 18 Feb 2020. The firm address in the FDA record is in Saint Cloud, MN. FDA put the recall in Class II. FDA event 89401 has 5 product records. The first FDA enforcement report date is 16 Feb 2022. The record gives the status Terminated, with the termination date 11 Jul 2022. 89 of the 916 device events that started in 2020 have more product records. The median device event of 2020 has 1 product record. The openFDA file has 916 device events that started in 2020 and 969 that started in 2019.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador,…
Recalling firm
Microbiologics Inc, Saint Cloud, MN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
5 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0585-2022 | KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299 | 1,975 units | Class II | Terminated | 16 Feb 2022 |
| Z-0586-2022 | LYFO-DISK Enterococcus faecalis derived from ATCC 51299 In Vitro Diagnostic Control microorganisms | 183 units | Class II | Terminated | 16 Feb 2022 |
| Z-0587-2022 | QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms | 10 units | Class II | Terminated | 16 Feb 2022 |
| Z-0588-2022 | QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms | 139 units | Class II | Terminated | 16 Feb 2022 |
| Z-0589-2022 | QC Sets and Panels, KWIK-STIK Blood Culture Identification (BCID) Verification Panel (Live Culture). KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms | 3 units | Class II | Terminated | 16 Feb 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2020
- 916
- Share of these events in Class II
- 93%
- Product records in this event
- 5Median of the firm: 1
- Events of the firm on the register
- 43
Other events of the firm
Official channels
- FDA enforcement report Z-0585-2022The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 89401. https://recallregister.com/fda/ii/89401/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador,…
- What quantity of product is in FDA event 89401?
- The event has 5 product records. The first record gives this quantity:
1,975 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2020 have the same class?
- 93% of the 916 device events that started in 2020 are in Class II.