Recall Register

Food, drugs, devices › Biomet, Inc.

Device recall: A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation o

FDA enforcement event 87095 · Class II · initiated 23 Dec 2020 · status Terminated · terminated 13 Apr 2023

DeviceVoluntary: Firm initiated1 product recordWarsaw, IN

Biomet, Inc. (Warsaw, IN) initiated this device recall on 23 Dec 2020; FDA lists 1 product record under event 87095, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 916 device enforcement events initiated in 2020 carry the same class.

Status: Terminated, terminated 13 Apr 2023. Initiation: voluntary: firm initiated. Biomet, Inc. has 84 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Biomet, Inc., Warsaw, IN.

Reason for recall

"The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill."

Distribution

"Nationwide distribution."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0882-2021A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.174 systemsClass IITerminated27 Jan 2021

Context

93%of 2020 device events are Class II916 events that year
1product record in this eventfirm median: 1
84events on this register for the firm

Other enforcement events for Biomet, Inc.

InitiatedProduct (first record)TypeClassRecords
1 Jul 2026Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade ADeviceClass II5
23 Sep 2025BIOMET ZipTight, Acute AC Joint Implant, Single Ziploop , Model/Catalog Number: 904834; Fixation System for AC Joint RepDeviceClass II1
9 Sep 2025Juggerknot Mini Soft Anchors, Model/Catalog Number: 912080DeviceClass II1
21 Oct 2024JuggerLoc/ToggleLoc 2.9mm Drill Guide, Model number 110027357DeviceClass II1
17 May 2024RingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number: 11-165206DeviceClass II1
13 Mar 2024Vanguard Knee System AS Tibial Bearing / DCM ArCom, 12 MM X 75 MM; Model Number 189082. The product is placed into a foiDeviceClass II1
2 Nov 2023Biolox Ceramic Option Head Zimmer 12/14 40mm -3, Item Number 00-8777-040-01DeviceClass II2
9 Oct 2023Compress Device Segmental Anchor Plug, 10 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previDeviceClass II19
6 Feb 2023G7 Acetabular System, Dual Mobility Acetabular Liner, 40 mm, Size D, Model Number 110024462DeviceClass II1
29 Dec 2022(1) Vanguard Knee System PS Open Box Femoral; Right; 60 mm. Intended for knee joint replacement prostheses. Item Number:DeviceClass II2

All 84 events for Biomet, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Nationwide distribution."

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 13 Apr 2023. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.