Recall Register

Device recall: The Bard Intra-abdominal Pressure Monitoring Device is composed of a tubing set used for infusing fluid into the…

Class II FDA enforcement event 84744. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Leaks or disconnect near the sampling port of the Intra Abdominal Pressure Monitoring Device which could lead to the potential of inaccurate readings. The measurement should be confirmed with other means, such as a new IAP system of the same kind or by choosing other diagnostic means such as but not limited to alternative systems, such as open IAP measurement options or exchanging the tubing…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

C.R. Bard Inc started this device recall on 13 Jan 2020. The firm address in the FDA record is in Covington, GA. FDA put the recall in Class II. FDA event 84744 has 1 product record. The first FDA enforcement report date is 19 Feb 2020. The record gives the status Terminated, with the termination date 12 Apr 2021. 324 of the 916 device events that started in 2020 have more product records. The median device event of 2020 has 1 product record. The openFDA file has 916 device events that started in 2020 and 969 that started in 2019.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Distributed Nationwide including Washington DC and Guaynabo PR; Distributed OUS to Belgium and Japan.

Recalling firm

C.R. Bard Inc, Covington, GA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 84744: product records
Recall numberProductQuantityClassStatusReport date
Z-1219-2020The Bard Intra-abdominal Pressure Monitoring Device is composed of a tubing set used for infusing fluid into the urinary bladder through the Foley catheter sampling port. It utilizes a clamping device to occlude the urinary drainage tubing to form a fluid column through which pressure is measured.1,300 totalClass IITerminated19 Feb 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
49

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84744. https://recallregister.com/fda/ii/84744/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Distributed Nationwide including Washington DC and Guaynabo PR; Distributed OUS to Belgium and Japan.
What quantity of product is in FDA event 84744?
The FDA record gives this quantity: 1,300 total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.