Recall Register

Food, drugs, devices › Physio-Control, Inc.

Device recall: LIFEPAK 15 MONITOR/DEFIBRILLATOR that was either manufactured with or received an upgrade kit that contained an affected

FDA enforcement event 84425 · Class II · initiated 20 Dec 2019 · status Terminated · terminated 4 Oct 2022

DeviceVoluntary: Firm initiated2 product recordsRedmond, WA

Physio-Control, Inc. (Redmond, WA) initiated this device recall on 20 Dec 2019; FDA lists 2 product records under event 84425, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 968 device enforcement events initiated in 2019 carry the same class.

Status: Terminated, terminated 4 Oct 2022. Initiation: voluntary: firm initiated. Physio-Control, Inc. has 27 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Physio-Control, Inc., Redmond, WA.

Reason for recall

"Monitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button."

Distribution

"Worldwide Distribution - U.S Nationwide: GA, PA, IN, CA, KY, FL, AZ, WA, NC, MD, TX, CO, VT, MI, MA, TN, OH, CT, NY, MO, VA, IA, MN, LA, WV, MT, IL, AR, UT, NJ, OK, NV, NH, MS, NM, WI, OR, ID, SC,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0816-2020LIFEPAK 15 MONITOR/DEFIBRILLATOR that was either manufactured with or received an upgrade kit that contained an affected keypad.29952Class IITerminated22 Jan 2020
Z-0817-2020LIFEPAK 15 MONITOR/DEFIBRILLATOR service kits705Class IITerminated22 Jan 2020

Context

88%of 2019 device events are Class II968 events that year
2product records in this eventfirm median: 1
27events on this register for the firm

Other enforcement events for Physio-Control, Inc.

InitiatedProduct (first record)TypeClassRecords
13 Apr 2026LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577DeviceClass II1
21 Jan 2026PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 007219026DeviceClass II1
22 Jul 2025LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 DeviceClass II3
3 Feb 2025LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.DeviceClass II1
21 Jan 2025LIFEPAK 35 ECG cable REF 11111-000041DeviceClass II1
21 Jan 2025LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo¿ RD rainbow Adt 8¿ SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-DeviceClass II1
30 Aug 2024LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99DeviceClass II1
9 Aug 2024STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP20 STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP15 DefibrillatiDeviceClass II1
6 Nov 2023AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098DeviceClass II1
29 Sep 2023LIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051DeviceClass II10

All 27 events for Physio-Control, Inc.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Monitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide Distribution - U.S Nationwide: GA, PA, IN, CA, KY, FL, AZ, WA, NC, MD, TX, CO, VT, MI, MA, TN, OH, CT, NY, MO, VA, IA, MN, LA, WV, MT, IL, AR, UT, NJ, OK, NV, NH, MS, NM, WI, OR, ID, SC,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 4 Oct 2022. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance